beautypg.com

Test port input (trace noise) – Atec Agilent-E5071C User Manual

Page 22

background image

22

Test port input (trace noise)

(continued)

Table 21.

Option 2D5/2K5/4D5/4K5

Description

Specifi cation

SPD

Trace noise magnitude

1, 3

300 kHz to 1 MHz (IFBW = 3 kHz)

1 MHz to 10 MHz (IFBW = 3 kHz)

10 MHz to 4.38 GHz (IFBW = 70 kHz)

4.38 GHz to 8.5 GHz (IFBW = 70 kHz)

8.5 GHz to 13.137 GHz (IFBW = 70 kHz)

13.137 GHz to 17 GHz (IFBW = 70 kHz)

17 GHz to 20 GHz (IFBW = 70 kHz)
(at maximum output power level
of sweep range)

0.006 dB rms

0.003 dB rms

0.004 dB rms

0.006 dB rms

0.009 dB rms

0.013 dB rms

0.023 dB rms

0.0009 dB rms

0.0005 dB rms

0.0010 dB rms

0.0012 dB rms

0.0024 dB rms

0.0040 dB rms

0.0065 dB rms

Trace noise phase

2, 3

300 kHz to 1 MHz (IFBW = 3 kHz)

1 MHz to 10 MHz (IFBW = 3 kHz)

10 MHz to 4.38 GHz (IFBW = 70 kHz)

4.38 GHz to 8.5 GHz (IFBW = 70 kHz)

8.5 GHz to 13.137 GHz (IFBW = 70 kHz)

13.137 GHz to 17 GHz (IFBW = 70 kHz)

17 GHz to 20 GHz (IFBW = 70 kHz)
(at maximum output power level
of sweep range)

0.040 deg rms

0.020 deg rms

0.035 deg rms

0.050 deg rms

0.064 deg rms

0.095 deg rms

0.165 deg rms

0.0120 deg rms

0.0025 deg rms

0.0075 deg rms

0.0150 deg rms

0.0250 deg rms

0.0320 deg rms

0.0520 deg rms

1. The specifi cation might not be met at the following frequencies: 406.25 kHz, 666.667 kHz,

722.222 kHz, 857.143 kHz, 928.571 kHz, 1.444444 MHz, 1.714286 MHz, 1.8 MHz, 1.857143 MHz,

1.95 MHz, 2.4375 MHz, 2.571429 MHz, 3.714286 MHz, 4.8 MHz, 5 MHz, 5.416667 MHz,

7.583333 MHz, 10 MHz, 10.833333 MHz, 12.5 MHz.

2. The specifi cation might not be met at 5 MHz or 50 MHz.