Measurement throughput summary, Cycle time for measurement completion – Atec Agilent-E5071C User Manual
Page 37
37
Measurement Throughput Summary
Cycle time for measurement completion
1, 2
(ms)
Table 36.
Option 230/235/240/245/260/265/280/285/430/435/440/445/
460/465/480/485
Sweep mode: Swept, analyzer display turned off with: DISP:ENAB OFF,
number of traces = 1, system error correction: OFF
Number of points
51
201
401
1601
Start 1 GHz, stop 1.2 GHz, 500 kHz IF bandwidth
Uncorrected
3.4
4.9
7.1
20
2-port cal
6.3
9.3
14
41
4-port cal
13
19
28
92
Start 1 GHz, stop 1.2 GHz, 100 kHz IF bandwidth
Uncorrected
3.7
5.2
7.3
20
2-port cal
6.6
10
15
42
4-port cal
13
21
32
94
Start 1 GHz, stop 1.2 GHz, 1 kHz IF bandwidth
Uncorrected
54
202
397
1563
2-port cal
107
401
792
3125
4-port cal
213
800
1581
6247
Start 100 kHz, stop 3 GHz, 500 kHz IF bandwidth
Uncorrected
8.7
11
12
25
2-port cal
17
21
24
48
4-port cal
34
42
49
102
Start 100 kHz, stop 3 GHz, 100 kHz IF bandwidth
Uncorrected
8.8
11
13
26
2-port cal
17
22
24
54
4-port cal
34
42
49
118
Start 100 kHz, stop 3 GHz, 1 kHz IF bandwidth
Uncorrected
56
206
404
1583
2-port cal
112
410
805
3163
4-port cal
222
818
1608
6324
1. Supplemental performance data.
2. Measured with fi rmware version A.09.30.