33

Table 35.

Analyzer environmental specifications and dimensions

Description

General characteristics

Operating environment

Temperature

+5 °C to +40 °C

Error-corrected temperature range

23 °C (±5 °C) with < 1 °C deviation from calibration temperature

Humidity

20% to 80% at wet bulb temperature < +29 °C (non-condensation)

Altitude

0 to 2,000 m (0 to 6561 feet)

Vibration

0.21 G maximum, 5 Hz to 500 Hz

Non-Operating storage environment

Temperature

–10 °C to +60 °C

Humidity

20% to 90% at wet bulb temperature < +40 °C (non-condensation)

Altitude

0 to 4572 m (0 to 15,000 feet)

Vibration

0.5 G maximum, 5 Hz to 500 Hz

Dimensions

See next page.

Weight (net)
Option 230/240/260/280 (2-port)
Option 235/245/265/285 (2-port)
Option 430/480/440/460 (4-port)
Option 435/445/465/485 (4-port)
Option 2D5/2K5 (2-port)
Option 4D5/4K5 (4-port)

18.2 kg
18.3 kg
19.9 kg
20.0 kg
19.8 kg
21.8 kg