4

Corrected System Performance with Type-N Connectors


Table 2. Corrected System Performance With Type-N Device Connectors, 85032F Calibration
Kit

Applies to the E8801A, E8802A, and E8803A analyzer, 85032F (Type-N, 50

) calibration kit, N6314A test port cable, and a full 2-port

calibration. Also applies to the following conditions:

IF bandwidth = 10 Hz

No averaging applied to data

Environmental temperature 25° ±5 °C, with < 1 °C deviation from calibration temperature

Isolation calibration not omitted

Specification (dB)

Description

300 kHz to

1.3 GHz

1.3 GHz to

3 GHz

3 GHz to

6 GHz

6GHz to

9 GHz

Directivity

49

46

40

38

Source Match

41

40

36

35

Load Match

49

45

39

37

Reflection Tracking

±0.011

±0.021

±0.032

±0.054

Transmission
Tracking

±0.012

±0.020

±0.055

±0.083

Transmission Uncertainty (Specifications)

Reflection Uncertainty (Specifications)