19
Table 20. Analyzer Environment and Dimensions
Description
Supplemental Information
General Environmental
RFI/EMI Susceptibility
Defined by CISPR Pub. 11, Group 1, Class A, and IEC
50082-1
ESD
Minimize using static-safe work procedures and an
antistatic bench mat
Dust
Minimize for optimum reliability
Operating Environment
Temperature
0 °C to +40 °C
Instrument powers up, phase locks, and displays no
error messages within this temperature range.
Error-Corrected Temperature Range
25°C ± 5°C
with less than 1°C deviation from calibration temp.
Humidity
5% to 95% at +40 °C
Altitude
0 to 4500 m (14,760 ft.)
Non-Operating Storage Environment
Temperature
-40 °C to +70 °C
Humidity
0% to 90% at +65 °C (non-condensing)
Altitude
0 to 15,240 m (50,000 ft.)
Cabinet Dimensions
Height
Width
Depth
Excluding front and rear
panel hardware and feet
223 mm
8.75 in
426 mm
16.75 in
427 mm
16.8 in
As shipped - includes front
panel connectors, rear
panel bumpers, and feet.
235 mm
9.25 in
435 mm
17.10 in
470 mm
18.5 in
As shipped plus handles
235 mm
9.25 in
458 mm
18 in
501 mm
19.70 in
As shipped plus rack-
mount flanges
235 mm
9.25 in
483 mm
19 in
470 mm
18.5 in
As shipped plus handles
and flanges
235 mm
9.25 in
483 mm
19 in
501 mm
19.70 in
Weight
Net
24 kg (54 lb), nominal
Shipping
32 kg (70 lb), nominal
Note: "Misc. Information" specifications are not included in this E8801/2/3A document.