20
Measurement Throughput Summary
Table 21. Typical Cycle Time
a,b
(ms)
Number of Points
101
201
401
1601
Start 1.8 GHz, Stop 2 GHz, 35 kHz IF bandwidth
Uncorrected,
1-port cal
7
10
16
52
2-Port cal
27
36
55
164
Start 300 kHz, Stop 3 GHz, 35 kHz IF bandwidth
Uncorrected,
1-port cal
48
54
64
104
2-Port cal
103
119
145
254
Start 300 kHz, Stop 9 GHz, 35 kHz IF bandwidth
Uncorrected,
1-port cal
51
57
64
103
2-Port cal
112
124
138
220
a Typical performance.
b Includes sweep time, retrace time and band-crossing time. Analyzer display turned off with
DISPLAY:ENABLE OFF. Add 21 ms for display on. Data for one trace (S11) measurement..
Table 22. Cycle Time vs. IF Bandwidth
a
Applies to the
Preset condition
(201 p
oints, correction off) except for the following changes:
•
CF = 1 GHz
•
Span = 100 MHz
•
Display off (add 21 ms for display on)
IF Bandwidth
(Hz)
Cycle Time (ms)
b
40,000
8
35,000
9
30,000
11
20,000
13
10,000
28
7000
36
5000
48
3000
72
1000
196
300
620
100
1875
30
8062
10
17877
a
Typical performance.
b
Cycle time includes sweep and retrace time.