beautypg.com

Atec Agilent-E8802A User Manual

Page 24

background image

20

Measurement Throughput Summary

Table 21. Typical Cycle Time

a,b

(ms)

Number of Points

101

201

401

1601

Start 1.8 GHz, Stop 2 GHz, 35 kHz IF bandwidth

Uncorrected,

1-port cal

7

10

16

52

2-Port cal

27

36

55

164

Start 300 kHz, Stop 3 GHz, 35 kHz IF bandwidth

Uncorrected,

1-port cal

48

54

64

104

2-Port cal

103

119

145

254

Start 300 kHz, Stop 9 GHz, 35 kHz IF bandwidth

Uncorrected,

1-port cal

51

57

64

103

2-Port cal

112

124

138

220

a Typical performance.

b Includes sweep time, retrace time and band-crossing time. Analyzer display turned off with

DISPLAY:ENABLE OFF. Add 21 ms for display on. Data for one trace (S11) measurement..

Table 22. Cycle Time vs. IF Bandwidth

a

Applies to the

Preset condition

(201 p

oints, correction off) except for the following changes:

CF = 1 GHz

Span = 100 MHz

Display off (add 21 ms for display on)

IF Bandwidth
(Hz)

Cycle Time (ms)

b

40,000

8

35,000

9

30,000

11

20,000

13

10,000

28

7000

36

5000

48

3000

72

1000

196

300

620

100

1875

30

8062

10

17877

a

Typical performance.