11
Table 10. Test Port Output Signal Purity
Description
Specification
Supplemental Information
Harmonics (2nd or 3rd)
at max output power (< 25 MHz)
< -25 dBc, typical
at max output power (25 MHz to 9 GHz)
< -25 dBc, characteristic
a
at 0 dBm output
< -35 dBc, typical
at -10 dBm output
< -38 dBc, typical, in power
range 0
Non-harmonic Spurious
at max output
-30 dBc, typical for offset freq>1kHz
at -10 dBm output
-50 dBc, typical for offset freq >1kHz
a
Typical below 25 MHz.
Test Port and Receiver Input Characteristics
Table 11. Test Port and Receiver Input Levels
Description
Specification
Supplemental Information
Maximum Test Port Input Level
Test Ports 1 and 2:
300 kHz to 25 MHz
+10 dBm
< 0.6 dB compression
25 MHz to 3 GHz
+10 dBm
< 0.4 dB compression
3 GHz to 6 GHz
+10 dBm
< 0.7 dB compression
6 GHz to 9 GHz
+5 dBm
< 0.7 dB compression
Damage Level
Test Port 1, 2
+30 dBm or ±30 VDC, typ.
R, A, B (Opt. 014)
+15 dBm or ±5 VDC, typ.
Coupler Thru (Opt. 014)
+33 dBm or ±0 VDC, typ.
Test Port Noise Floor
a
300 kHz to 25 MHz
b
10 Hz IF Bandwidth
-115 dBm
1 kHz IF Bandwidth
-95 dBm
25 MHz to 3 GHz
b
10 Hz IF Bandwidth
-118 dBm
1 kHz IF Bandwidth
-98 dBm
3 GHz to 9 GHz
10 Hz IF Bandwidth
≤
-108 dBm
1 kHz IF Bandwidth
≤
-88 dBm