11

Table 10. Test Port Output Signal Purity

Description

Specification

Supplemental Information

Harmonics (2nd or 3rd)

at max output power (< 25 MHz)

< -25 dBc, typical

at max output power (25 MHz to 9 GHz)

< -25 dBc, characteristic

a

at 0 dBm output

< -35 dBc, typical

at -10 dBm output

< -38 dBc, typical, in power

range 0

Non-harmonic Spurious

at max output

-30 dBc, typical for offset freq>1kHz

at -10 dBm output

-50 dBc, typical for offset freq >1kHz

a

Typical below 25 MHz.


Test Port and Receiver Input Characteristics


Table 11. Test Port and Receiver Input Levels

Description

Specification

Supplemental Information

Maximum Test Port Input Level

Test Ports 1 and 2:

300 kHz to 25 MHz

+10 dBm

< 0.6 dB compression

25 MHz to 3 GHz

+10 dBm

< 0.4 dB compression

3 GHz to 6 GHz

+10 dBm

< 0.7 dB compression

6 GHz to 9 GHz

+5 dBm

< 0.7 dB compression

Damage Level

Test Port 1, 2

+30 dBm or ±30 VDC, typ.

R, A, B (Opt. 014)

+15 dBm or ±5 VDC, typ.

Coupler Thru (Opt. 014)

+33 dBm or ±0 VDC, typ.

Test Port Noise Floor

a

300 kHz to 25 MHz

b

10 Hz IF Bandwidth

-115 dBm

1 kHz IF Bandwidth

-95 dBm

25 MHz to 3 GHz

b

10 Hz IF Bandwidth

-118 dBm

1 kHz IF Bandwidth

-98 dBm

3 GHz to 9 GHz

10 Hz IF Bandwidth

-108 dBm

1 kHz IF Bandwidth

-88 dBm