beautypg.com

Atec Agilent-E8802A User Manual

Page 14

background image

10

Table 9. Test Port Output Power

a

Description

Specification

Supplemental Information

Level Accuracy:

300 kHz to 6 GHz

±1.0 dB

Variation from 0 dBm in power range 0

6 GHz to 9 GHz

±2.0 dB

±1.5dB below 10 MHz

Level Linearity:

Variation from 0 dBm in power range 0

300 kHz to 9 GHz

±0.3 dB

-15 to +5 dBm

300 kHz to 1 MHz

±1.0 dB

+5 to +10 dBm

1 MHz to 6 GHz

±0.5 dB

+5 to +10 dBm

6 GHz to 9 GHz

±0.5 dB

+5 to +7 dBm

Range

b:

300 kHz to 6 GHz

-15 to +10 dBm

6 GHz to 9 GHz

-15 to +7 dBm

Range

b:

(Option 1E1):

300 kHz to 6 GHz

-85 to +10 dBm

6 GHz to 9 GHz

-85 to +7 dBm

Sweep Range

300 kHz to 6 GHz

25 dB

6 GHz to 9 GHz

22 dB

Level Resolution

0.01 dB

a

Source output performance on port 1 only. Port 2 output performance is typical.

b

Power to which the source can be set and phase lock is assured.