beautypg.com

Atec Agilent-E8802A User Manual

Page 10

background image

6

Corrected System Performance with 3.5 mm Connectors

Table 4. Corrected System Performance With 3.5 mm Device Connector Type, 85033E
Calibration Kit

Applies to the E8801A, E8802A, and E8803A analyzer, 85033E (3.5 mm, 50

) calibration kit, N6314A test port cable, and a full 2-port

calibration. Also applies to the following conditions:

IF bandwidth = 10 Hz

No averaging applied to data

Environmental temperature 25° ±5 °C, with < 1 °C deviation from calibration temperature

Isolation calibration not omitted

Specification (dB)

Description

300 kHz to

1.3 GHz

1.3 GHz to

3 GHz

3 GHz to

6 GHz

6 GHz to

9 GHz

Directivity

46

44

38

38

Source Match

43

40

37

36

Load Match

46

44

38

38

Reflection Tracking

±0.006

±0.007

±0.009

±0.010

Transmission
Tracking

±0.012

±0.021

±0.057

±0.075

Transmission Uncertainty (Specifications)

Reflection Uncertainty (Specifications)