beautypg.com

Atec Agilent-E8802A User Manual

Page 13

background image

9

Table 7. Uncorrected Instrument Performance

Specification (dB)

Description

300 kHz to

1 MHz

1MHz to

1.3 GHz

1.3 GHz to

3 GHz

3 GHz to

6 GHz

6 GHz to

9 GHz

Directivity

30

33

27

20

13

Source Match

18

18

16

11

8

Load Match

20

20

17

13.5

13

Reflection
Tracking

±1.5

±1.5

±1.5

±2.5

±3.0

Transmission
Tracking

±1.5

±1.5

±1.5

±2.5

±3.0

Test Port Output Characteristics (Source)

Table 8. Test Port Output Frequency

Description

Specification

Supplemental Information

Range:

E8801A

300 kHz to 3.0 GHz

E8802A

300 kHz to 6.0 GHz

E8803A

300 kHz to 9.0 GHz

Resolution:

1 Hz

Source Stability

±1 ppm, 0°C to 40 °C, typical

±1ppm/year maximum

Source Stability

(Option 1E5)

±0.05 ppm, 0° to 70 °C, typical

±0.1 ppm/year maximum

CW Accuracy

±3 ppm

CW Accuracy

(Option 1E5)

±1 ppm