Table 15. Test Port Input (Group Delay)

a

Description

Specification

Supplemental Information

Aperture (selectable)

(frequency span)/(number of
points -1)

Maximum Aperture

20% of frequency span

Range

0.5 x (1/minimum aperture)

Maximum Delay

Limited to measuring no more
than 180° of phase change within
the minimum aperture.)

Accuracy

See graph below. Char.

The following graph shows group delay accuracy with type-N full 2-port calibration and a 10 Hz IF
bandwidth. Insertion loss is assumed to be < 2 dB and electrical length to be ten meters.

In general, the following formula can be used to determine the accuracy, in seconds, of specific group
delay measurement:

±Phase Accuracy (deg)/[360 × Aperture (Hz)]

Depending on the aperture and device length, the phase accuracy used is either incremental phase
accuracy or worst case phase accuracy.

a

Group delay is computed by measuring the phase change within a specified frequency step (determined by the frequency span and the

number of points per sweep).

General Information


Table 16. System Bandwidths

Description

Specification

Supplemental Information

IF Bandwidth Settings

Range

1 Hz to 40 kHz

in a 1, 2, 3, 5, 7,10 sequence up to 30 kHz, 35 kHz, 40kHz,
nominal