8

Table 6. Corrected System Performance With 7-16 Device Connector Type, 85038A Calibration
Kit

Applies to the E8801A, E8802A, and E8803A analyzer, 85038A (7-16, 50

) calibration kit, N6314A test port cable, and a full 2-port

calibration. Also applies to the following conditions:

IF bandwidth = 10 Hz

No averaging applied to data

Environmental temperature 25° ±5 °C, with < 1 °C deviation from calibration temperature

Isolation calibration not omitted

Specification (dB)

Description

300 kHz to

1.3 GHz

1.3 GHz to

3 GHz

3 to

6 GHz

6 to

9 GHz

a

Directivity

40

40

36

36

Source Match

37

37

34

34

Load Match

39

39

35

35

Reflection Tracking

±0.089

±0.089

±0.115

±0.115

Transmission
Tracking

±0.024

±0.033

±0.082

±0.103

Transmission Uncertainty (Specifications)