beautypg.com

Atec Agilent-E8802A User Manual

Page 11

background image

7

Table 5. Corrected System Performance With 3.5 mm Device Connector Type, 85093C
Electronic Calibration Module

Applies to the E8801A, E8802A, and E8803A analyzer, 85093C (3.5 mm, 50

) electronic calibration (ECal) module, N6314A test port cable,

and a full 2-port calibration. Also applies to the following conditions:

IF bandwidth = 10 Hz

No averaging applied to data

Environmental temperature 25° ±5 °C, with < 1 °C deviation from calibration temperature

Isolation calibration not omitted

Specification (dB)

Description

300 kHz to

1.3 GHz

1.3 GHz to

3 GHz

3 GHz to

6 GHz

6 GHz to

9 GHz

a

Directivity

52

52

51

47

Source Match

44

44

39

34

Load Match

47

47

44

40

Reflection Tracking

±0.030

±0.040

±0.050

±0.070

Transmission
Tracking

±0.039

±0.049

±0.068

±0.117

Transmission Uncertainty (Specifications)

Reflection Uncertainty (Specifications)