beautypg.com

Table 32. cycle time vs. if bandwidth - typical – Atec Agilent-N5227A User Manual

Page 40

background image

40

Table 31. Typical Cycle Time (ms) for Full-Span Measurement Completion

10 MHz to 67 GHz

Number of Points

IF Bandwidth

201

401

1601

16001

32001

600 kHz

Uncorrected

55.5

72.7

94.5

182

248

2-Port cal

117

152

195

374

509

10 kHz

Uncorrected

89.1

153

519

2219

4223

2-Port cal

185

313

1042

4448

8462

1 kHz

Uncorrected

255

483

1834

17716

35172

2-Port cal

515

972

3675

35444

70375

Table 32. Cycle Time vs. IF Bandwidth - Typical

Applies to the Preset condition (201 points, correction off) except for the following changes:

CF = 10 GHz

Span = 100 MHz

Display off (add 21 ms for display on)

Cycle time includes sweep and retrace time.

Description

N5227A

IF Bandwidth (Hz)

Cycle Time (ms)

Trace Noise Magnitude (dB rms)

600,000

6.3

0.0044

100,000

7

0.0021

30,000

10.2

0.0011

10,000

29.7

0.0007

3,000

71.9

0.0006

1,000

223

0.0004

300

641

0.0004

100

1825

0.0003

30

5981

0.0003

10

17834

0.0003

3

59273

0.0003