Test Port Output

Table 6. Frequency Information, All Options

Description

Specification

Typical

N5227A Frequency Range

10 MHz to 67 GHz

67 GHz to 70 GHz

Frequency Resolution

1 Hz

--

Frequency Accuracy

+/- 1 ppm

--

Frequency Stability

--

+/-0.05 ppm, -10° to 70° C1

+/-0.1 ppm/yr maximum2






+/-0.1 ppm/yr maximum2

1

Assumes no variation in time.

2

Assumes no variation in temperature.

Table 7a. Maximum Leveled Power (dBm) - Specification

Description

Option 200, 400

Option 201, 401

Option 219, 419

Port 1,

Port 3

Port 2,
Port 4

Port 1,

Port 3

Port 2,
Port 4

Port 1,

Port 3

Port 2,
Port 4

10 MHz to 50 MHz

12

12

12

12

11

11

50 MHz to 2 GHz

13

13

13

13

13

13

2 GHz to 3.2 GHz

10

13

10

13

9

13

3.2 GHz to 10 GHz

13

13

13

13

11

12

10 GHz to 13.5 GHz

12

12

11

11

9

9

13.5 GHz to 16 GHz

12

12

12

12

10

10

16 GHz to 19 GHz

11

11

10

10

8

8

19 GHz to 24 GHz

11

11

11

11

8

8

24 GHz to 26.5 GHz

11

11

11

11

7

7

26.5 GHz to 30 GHz

10

10

10

10

7

7

30 GHz to 32 GHz

9

9

7

7

5

5

32 GHz to 35 GHz

10

10

9

9

6

6

35 GHz to 40 GHz

5

5

5

5

1

1

40 GHz to 50 GHz

11

11

10

10

6

6

50 GHz to 64 GHz

11

11

10

10

5

5

64 GHz to 67 GHz

11

11

10

10

4

4