Table 19. N5227A Trace Noise Magnitude (dB rms)
Ratioed measurement, nominal power at test port.
Description
Specification
Typical
1 kHz IFBW
1 kHz IFBW
100 kHz IFBW
600 kHz IFBW
10 MHz to 50 MHz
0.05
0.0177
0.173
0.416
50 MHz to 100 MHz
0.004
0.0012
0.012
0.029
100 MHz to 500 MHz
0.002
0.0006
0.006
0.014
500 MHz to 1 GHz
0.002
0.0004
0.003
0.006
1 GHz to 26.5 GHz
0.002
0.0005
0.002
0.005
26.5 GHz to 50 GHz
0.003
0.0006
0.005
0.012
50 GHz to 67 GHz
0.003
0.0007
0.006
0.013
67 GHz to 70 GHz
--
0.0010
0.007
0.016
Table 20. N5227A Trace Noise Phase (deg rms)
Ratioed measurement, nominal power at test port.
Description
Specification
Typical
1 kHz IFBW
1 kHz IFBW
100 kHz IFBW
600 kHz IFBW
10 MHz to 50 MHz
0.400
0.1228
1.205
2.928
50 MHz to 100 MHz
0.020
0.0083
0.080
0.196
100 MHz to 500 MHz
0.020
0.0040
0.037
0.097
500 MHz to 1 GHz
0.020
0.0017
0.015
0.037
1 GHz to 26.5 GHz
0.020
0.0075
0.015
0.031
26.5 GHz to 43.5 GHz
0.030
0.0125
0.040
0.091
43.5 GHz to 50 GHz
0.035
0.0149
0.040
0.092
50 GHz to 67 GHz
0.045
0.0200
0.048
0.110
67 GHz to 70 GHz
--
0.0213
0.050
0.119
Table 21. Reference Level Magnitude, All Models and Options - Specification
Description
Magnitude (dB)
Phase (degrees)
Range
+/- 500
+/- 500
Resolution
0.001
0.01