Table 28. (Continued) Rear Panel Information, All Options
Description
Typical Performance
External IF Inputs
Function
Allows use of external IF signals from remote mixers, bypassing the PNA's first
converters
Connectors
SMA (female); A, B, C, D, R (4-port); A, B, R1, R2 (2-port)
Input Frequency
Normal IF path
Narrowband IF path
RF < 53 MHz: IF = 826.446 KHz
RF >= 53 MHz: IF = 7.438 MHz
IF = 10.70 MHz
Input Impedance
50
RF Damage Level
+23 dBm
DC Damage Level
5.5 VDC
0.1 dB Compression Point
Normal IF path
Narrowband IF path
-9.0 dBm at 7.438 MHz
-17 dBm at 10.70 MHz
Pulse Inputs (IF Gates)
Function
Internal receiver gates used for point-in-pulse and pulse-profile measurements
Connectors
15-pin mini D-sub
Input Impedance
1 K Ohm
Minimum Pulse Width,
Source Modulators
33 ns
Minimum Pulse Width,
Receiver Gates
20 ns
DC Damage Level
5.5 VDC
Drive Voltage
0 V (off), +3.3 V (on), nominal
RF Pulse Modulator Input (Source Modulator)
On/Off Ratio
10 MHz to 3.2 GHz
-64
3.2 GHz to 67 GHz
-80
Pulse Period
Minimum
33 ns
Maximum
70 s