N5227a corrected system performance, all options, Table 4a. n5227a with 85058b calibration kit – Atec Agilent-N5227A User Manual
Page 10
10
N5227A Corrected System Performance, All Options
Note: For any Sii reflection measurement:
Sjj = 0.
For any Sij transmission measurement:
Sji = Sij when Sij 1
Sji = 1/Sij when Sij 1
Skk = 0 for all k
Applies to the N5227A Option 200, 201, 219, 400, 401, or 419 analyzers, N4697F flexible test port cable set, and a full 2-
port calibration. Also applies to the following condition:
Environmental temperature 23° ±3 °C, with < 1 °C deviation from calibration temperature
Table 4a. N5227A with 85058B Calibration Kit
Description
Specification (dB)
10 MHz
to
50 MHz
50 MHz
to
2 GHz
2 GHz
to
10 GHz
10 GHz
to
20 GHz
20 GHz
to
35 GHz
35 GHz
to
50 GHz
50 GHz
to
60 GHz
60 GHz
to
67 GHz
Directivity
35
35
38
38
37
37
34
34
Source Match
34
34
40
40
41
42
40
40
Load Match
34
35
37
37
36
36
33
33
Reflection Tracking
Mag
0.019
0.019
0.033
0.033
0.033
0.020
0.030
0.030
Phase (°)
0.125
0.125
0.218
0.218
0.218
0.132
0.198
0.198
Transmission Tracking
Mag
0.159
0.128
0.099
0.094
0.100
0.093
0.121
0.137
Phase (°)
1.047
0.845
0.655
0.619
0.663
0.616
0.801
0.903