N5227A Corrected System Performance, All Options

Note: For any Sii reflection measurement:

Sjj = 0.

For any Sij transmission measurement:

Sji = Sij when Sij 1

Sji = 1/Sij when Sij 1

Skk = 0 for all k

Applies to the N5227A Option 200, 201, 219, 400, 401, or 419 analyzers, N4697F flexible test port cable set, and a full 2-
port calibration. Also applies to the following condition:

Environmental temperature 23° ±3 °C, with < 1 °C deviation from calibration temperature

Table 4a. N5227A with 85058B Calibration Kit

Description

Specification (dB)

10 MHz

to

50 MHz

50 MHz

to

2 GHz

2 GHz

to

10 GHz

10 GHz

to

20 GHz

20 GHz

to

35 GHz

35 GHz

to

50 GHz

50 GHz

to

60 GHz

60 GHz

to

67 GHz

Directivity

35

35

38

38

37

37

34

34

Source Match

34

34

40

40

41

42

40

40

Load Match

34

35

37

37

36

36

33

33

Reflection Tracking

Mag

0.019

0.019

0.033

0.033

0.033

0.020

0.030

0.030

Phase (°)

0.125

0.125

0.218

0.218

0.218

0.132

0.198

0.198

Transmission Tracking

Mag

0.159

0.128

0.099

0.094

0.100

0.093

0.121

0.137

Phase (°)

1.047

0.845

0.655

0.619

0.663

0.616

0.801

0.903