beautypg.com

Atec Agilent-85133F User Manual

Page 85

background image

Installing the System

2

Installation and User’s Guide

85

3 Connect the triaxial

cables from the
E5260A/70B SMUs to the
bias networks.

a Connect the triaxial cable from the

E5260A/70B medium power SMU to the
DC FORCE and DC SENSE connectors on
the port 1 bias network.

b Connect the triaxial cable from the

E5260A/70B high power SMU to the DC
FORCE and DC SENSE connectors on the
port 2 bias network.

The medium power SMU is the Agilent
E5281A.

The high power SMU is the Agilent
E5280A.

The port 1 bias network is the 11612V
K11.

The port 2 bias network is the 11612V
K21.

4 Connect and route the

ground cable from
E5260A/70B GNDU to the
port 2 bias network.

a From the front of the system, connect one

end of the ground triaxial cable to the
GNDU connector on the E5260A/70B front
panel.

b From the front of the system, route the

cable through the upper feedthrough panel.

c From the rear of the system, route the

ground triaxial cable through the port 2 hole
in the lower feedthrough panel.

d Connect the ground triaxial cable to the

port 2 bias network GNDU connector.

The ground triaxial cable model number
is 16493L Option 002.

Leave the GNDU connector on the port
1 bias network open.

5 Connect the Agilent

E8364B test ports to the
bias networks.

a Connect one end of the port 1 test port

cable to the Agilent E8364B test port 1.

b Connect the other end of the test port 1 test

port cable to the 2.4 mm female-to-male
adapter.

c Connect the 2.4 mm female-to-male

adapter to the RF IN connector on the
11612V Option K11 bias network.

d Connect one end of the port 2 test port

cable to the Agilent E8364B test port 2.

e Connect the other end of the test port 2 test

port cable to the 2.4 mm male-to-male
adapter.

f

Connect the other end of the 2.4 mm
male-to-male adapter to the RF IN
connector of the 11612V K21 bias network.

The test port cables model number is
85133F.

The 2.4 mm female-to-male adapter
model number is 11900C. Without this
adapter, the test port cable will not
properly mate with the bias network
RF IN connector.

The 2.4 mm male-to-male adapter
model number is 11900A.

To connect the bias networks (continued)

Step

Action

Notes