beautypg.com

Performance characteristics and specifications, Table 15 supplemental system characteristics, Interference standards – Atec Agilent-85133F User Manual

Page 61: Table 15. supplemental system characteristics

background image

Introducing the Agilent 85225F Performance Modeling System

1

Installation and User’s Guide

61

Performance Characteristics and Specifications

Supplemental characteristics are not specifications, but are provided in

Table 15

for your convenience.

Interference Standards

The IEC/EN 61326- 1 and CISPR Publication 11 standards define the RFI
and EMI susceptibility of the performance modeling system.

Performance Modeling System Performance Specifications

The Agilent 85225F performance modeling system adheres to the
performance specifications of an Agilent E8364B PNA Series vector
network analyzer. Refer to

Appendix G

, “Network Analyzer Performance

Specification Summary,” starting on page 141.

Table 15 Supplemental System Characteristics

Characteristic

Value

Line voltage

115 volts nominal (90 volts to 132 volts) or 220 volts nominal
(210 volts to 250 volts)

Line frequency

48 Hz to 66 Hz

Circuit breaker amperage rating

6 amps (115 volts), 3.5 amps (220 volts)

Rack weight capacity

818 kilograms (1800 pounds) maximum loaded

Rack external dimensions

1620 mm high

×

600 mm wide

×

905 mm deep

Rack footprint (top view)