Enhancing Measurement Accuracy
A
Installation and User’s Guide
115
Performing a Coaxial System Measurement Calibration
Complete the following steps to perform a coaxial system measurement
calibration and increase the accuracy of your device measurements. Modify
these instructions as necessary to perform an in- fixture or on wafer
measurement calibration.
To perform the system measurement calibration
Step
Action
Note
1 Preset the network
analyzer.
On the network analyzer, press Preset.
2 Open the Calibration
Wizard and choose a
guided calibration.
a In the network analyzer’s Calibration menu,
select Calibration Wizard... .
b In the Calibration Wizard dialog box, select
Use Mechanical Stds... in the Guided
Calibrations group box.
3 Choose the DUT
connector types.
a In the Guided Calibration: Select DUT
Connectors dialog box, select the
appropriate DUT connector type for the
connections the Port 1 and Port 2 bias
networks.
b Click Next.
4 Choose the calibration kit. a In the Guided Calibration: Select Version 2
Cal Kits dialog box, select the appropriate
calibration kit (for example 85056D) for the
Port 1 and Port 2 bias network.
b Click Next.