To verify the shipment – Atec Agilent-85133F User Manual
Page 68
68
Installation and User’s Guide
2
Installing the System
To verify the shipment
Use
“System Receiving Checklist and Replaceable Parts” to:
• confirm the completeness of the shipment
• provide component part and model numbers required to order
replacement parts
All replacement items are available from Agilent Technologies. Part
numbers for replacement instrument subassemblies are listed in their
individual service manuals.
The majority of the system components are shipped preconfigured in the
system rack cabinet.
The PGUs (pulse generator units) and GNDU (active ground unit) are
factory- installed in the 41501B PGU expander.
With the exception of the bias networks, semi- rigid and SMU triaxial
cables, all other cables are connected at the factory.
N O T E
Other cables and accessories are shipped inside the rack-mounted storage drawer.
To verify the shipment
Step
Action
Notes
1 Verify that the serial
numbers on the rear panel
of the system instruments
match the serial numbers
listed in the shipping
documentation.
Compare the serial numbers listed in the
shipping documents with the serial numbers
on the instrument’s rear panel serial number
labels.
•
If an instrument serial number does not
match the shipping document, report
mismatched serial number to your
Agilent Technologies sales
representative.
•
If all instrument serial numbers match
the shipping documents, continue to
step 2
.
•
For a list of support contacts, see