Enhancing Measurement Accuracy
A
Installation and User’s Guide
113
5 Always perform a
measurement calibration
on the network analyzer
prior to making device
measurements.
a Prior to making device measurements,
Coaxial System Measurement
Calibration"
b Ensure that the calibration kit elements are
seated correctly and firmly in the test set
cable connectors.
•
See
for more information.
•
Use the Agilent 85056A 2.4 mm
precision calibration kit.
•
Use the Agilent 85056D 2.4 mm
economy calibration kit.
•
Use the Agilent 85033E 3.5 mm
economy calibration kit.
•
Use the Agilent 85052D 3.5 mm
economy calibration kit.
•
Torque measurement connections
using the torque wrench provided in
the calibration kit.
6 Calibrate the system
components at regular
intervals.
a Every 6 to 12 months:
•
Perform the required Agilent E8364B
PNA Series vector network analyzer
performance verification tests.
•
Perform the required Agilent 4156C
precision semiconductor parameter
analyzer performance verification tests.
•
Perform the required Agilent 4248A
precision LCR meter performance
verification tests.
•
Perform the required Agilent E5250A low
leakage switch mainframe performance
verification tests.
To enhance measurement accuracy (continued)
Step
Action
Notes