beautypg.com

Atec Agilent-8510C User Manual

Page 7

background image

7

Agilent 85107B-005, 007

45 MHz to 50 GHz

The following specifications describe the system
performance for the 8510C network analyzer in the
85107B-005 (adds step attenuators and bias tees to the
8517B test set) and 85107B-007 (adds high power and
high dynamic range to the 8517B test set) configuration
for 50 GHz measurements. Specifications not shown are
the same as those given for the 85107B-005. The system
hardware includes the following:

Test set: 8517B-027, 8517B-003 S-parameter test set
RF source: 83651B synthesized sweeper
Calibration kit: 85056A 2.4mm calibration kit
Calibration technique: Full two-port calibration with
sliding loads

Measurement uncertainty

Transmission measurements

Agilent 85106D

33 GHz to 110 GHz

The following specifications describe the system
performance for the 8510C network analyzer in the
85106D configuration for measurements from 33 to
110 GHz in four waveguide bands. For complete
specifications, refer to the 85106D system data sheet.
The system hardware includes the following:

Test set (two each): 85104A series millimeter-wave test
set modules
RF sources (two): 83621B synthesized sweepers
Calibration kit: 11644A series waveguide
calibration kit
Calibration technique: TRL two-port calibration

Dynamic range

(for transmission measurements)

Frequency range (GHz)

0.045–2 2–20

20–40

40–50

Maximum power
measured at port 2

+17 dBm

+13 dBm

+6 dBm –4 dBm

Reference power
at port 1 (nominal)

+5 dBm

+2 dBm –5 dBm –16 dBm

Minimum power
measured at port 2

–71 dBm

–92 dBm –89 dBm –91 dBm

Receiver dynamic range 88 dB

105 dB

95 dB 87 dB

System dynamic range 76 dB

94 dB 84 dB 75 dB

Dynamic range

(for transmission measurements)

Frequency range (GHz)
33–50

40–60

50–75

75–110

Maximum power
measured at port 2

+12 dBm +10 dBm +10 dBm

0 dBm

Reference power
at port 1 (nominal)

0 dBm

0 dBm

0 dBm

–3 dBm

Minimum power
measured at port 2

–87 dBm –87 dBm

–75 dBm

–79 dBm

Receiver dynamic range 99 dB

97 dB

85 dB

79 dB

System dynamic range

87 dB

87 dB

75 dB

76 dB

System dynamic range

System dynamic range