beautypg.com

Atec Agilent-8510C User Manual

Page 24

background image

24

Electronic calibration

Electronic calibration (ECal) is a precision, single connec-
tion, one- or two-port calibration technique that uses fully
traceable and verifiable electronic impedance standards.
ECal replaces the traditional calibration technique that
uses mechanical standards. ECal requires fewer connec-
tions. It removes the intensive operator interaction, which
is prone to errors. A full two-port calibration can be
accomplished with a single connection of the ECal
module and minimal operator interaction. This results
in a faster and more repeatable calibration. Calibrations
for non-insertable devices are equally convenient and
straight forward.

An 85097A PC interface module with control software
must be ordered to drive all ECal modules. The
85097A requires a customer-supplied PC meeting
the following minimum requirements:

• Windows 95

®

, Windows NT 4.0

®

or later operating

system

• 486 or later CPU
• 32 MB of RAM
• 10 MB available hard-disk space
• one of the following GPIB interface cards:

82340A/B, 82341C/D, National AT-GPIB/TNT,
National AT-GPIB/TNT (plug&play) or National
PCI-GPIB

ECal modules are controlled manually or automatically
via the 85097A PC interface module with control soft-
ware. The 85097A consists of a PC interface module, con-
trol software and a power supply.

The Agilent 85090 family of RF ECal modules provides
calibration across the complete frequency range of the
8753C/D/E vector network analyzers. The 85060 family of
microwave ECal modules provides calibrations through
26.5 GHz for the 8510B/C and 8719/20/22C/D vector net-
work analyzers.

ECal modules and available options

A, B

Connector type

1

Frequency range

ECal module

Available options

model number

7 mm

30 kHz to 6 GHz

2

85091A

1BN, 1BP, 910, UK6

7 mm

1 GHz to 18 GHz

85060B

001, 1BN, 1BP, 910, UK6

Type-N

30 kHz to 6 GHz

2

85092A

00F, 00M, 00A, 1BN, 1BP, 910, UK6

Type-N

1 GHz to 18 GHz

85064B

001, 00F, 00M, 00A, 1BN, 1BP, 910,

UK6

3.5 mm

30 kHz to 6 GHz

2

85093A

00F, 00M, 00A, 1BN, 1BP, 910, UK6

3.5 mm

1 GHz to 26.5 GHz

85062B

001,00F, 00M, 00A, 1BN, 1BP, 910, UK6

Options

Option

Description

001

Adds a 30 kHz to 6 GHz RF module

2

00F

Replace f/m connectors on ECal module(s) with f/f connectors

00M

Replace f/m connectors on ECal module(s) with m/m connectors

00A

Adds male-to-male and female-to-female adapters (also adds a 5/16” 90 N-cm (8 in-lb)
torque wrench to 3.5 mm modules)

1BN

Mil-STD 45662 calibration certificate

1BP

Mil-STD 45662 calibration certificate with measured data

910

Add an extra operating and service manual

UK6

Commercial calibration certificate with measured data

1.

For ECal modules with sexed (m-f) connectors, the standard modules have one female and one male connector.

2.

RF ECal modules are specified to operate from 300 kHz to 6 GHz, with typical performance down to 30 kHz.

A. 85060 series modules cover a frequency range of 1 GHz to either 18 or 26.5 GHz. The upper frequency is limited by

the connector cutoff frequency. Each module is supplied with a torque wrench and foam-padded wood storage box.

B. 85090 series modules cover a frequency range of 30 kHz to 6 GHz. Each module is supplied with a torque wrench and

foam-padded wood storage box.