System performance – Atec Agilent-8510C User Manual
Page 4

4
Agilent 8510E-005
45 MHz to 20 GHz
The following specifications describe the system
performance for the 8510C network analyzer in the
8510E-005 (replaces 85052D fixed loads with 85052B
sliding loads calibration kit) configuration. The system
hardware includes the following:
Test set: 8514B S-parameter test set
RF source: 83621B synthesized sweeper
Calibration kit: 85052B 3.5mm calibration kit
Calibration technique: Full two-port calibration with
sliding loads.
Dynamic range
(for transmission measurements)
Frequency range (GHz)
0.045–2
2–8
8–20
Maximum power
measured at port 2
+20 dBm
+11 dBm
+10 dBm
Reference power
at port 1 (nominal)
+2 dBm
–2 dBm
–6 dBm
Minimum power
measured at port 2
–66 dBm
–95 dBm
–95 dBm
Receiver dynamic range
86 dB
106 dB
105 dB
System dynamic range
68 dB
93 dB
89 dB
Measurement uncertainty
Frequency range (GHz)
RESIDUAL
0.045–2 2–8
8–20
Directivity
48 dB
44 dB
44 dB
Source match
40 dB
33 dB
31 dB
Load match
48 dB
44 dB
44 dB
Reflection tracking
±0.003 dB
±0.003 dB
±0.006 dB
Transmission tracking
±0.017 dB
±0.044 dB
±0.084 dB
Crosstalk
89 dB
115 dB
110 dB
Frequency range (GHz)
RAW (Typical)
0.045–2
2–8
8–20
Directivity
23 dB
23 dB
14 dB
Source match
17 dB
15 dB
11 dB
Load match
17 dB
15 dB
11 dB
Reflection measurements
Transmission measurements
System dynamic range
Magnitude Phase
Magnitude Phase
System performance