System performance – Atec Agilent-8510C User Manual

Page 4

background image

4

Agilent 8510E-005

45 MHz to 20 GHz

The following specifications describe the system
performance for the 8510C network analyzer in the
8510E-005 (replaces 85052D fixed loads with 85052B
sliding loads calibration kit) configuration. The system
hardware includes the following:

Test set: 8514B S-parameter test set
RF source: 83621B synthesized sweeper
Calibration kit: 85052B 3.5mm calibration kit
Calibration technique: Full two-port calibration with
sliding loads.

Dynamic range

(for transmission measurements)

Frequency range (GHz)
0.045–2

2–8

8–20

Maximum power
measured at port 2

+20 dBm

+11 dBm

+10 dBm

Reference power
at port 1 (nominal)

+2 dBm

–2 dBm

–6 dBm

Minimum power
measured at port 2

–66 dBm

–95 dBm

–95 dBm

Receiver dynamic range

86 dB

106 dB

105 dB

System dynamic range

68 dB

93 dB

89 dB

Measurement uncertainty

Frequency range (GHz)

RESIDUAL

0.045–2 2–8

8–20

Directivity

48 dB

44 dB

44 dB

Source match

40 dB

33 dB

31 dB

Load match

48 dB

44 dB

44 dB

Reflection tracking

±0.003 dB

±0.003 dB

±0.006 dB

Transmission tracking

±0.017 dB

±0.044 dB

±0.084 dB

Crosstalk

89 dB

115 dB

110 dB

Frequency range (GHz)

RAW (Typical)

0.045–2

2–8

8–20

Directivity

23 dB

23 dB

14 dB

Source match

17 dB

15 dB

11 dB

Load match

17 dB

15 dB

11 dB

Reflection measurements

Transmission measurements

System dynamic range

Magnitude Phase

Magnitude Phase

System performance