beautypg.com

Agilent 8510sx – Atec Agilent-8510C User Manual

Page 5

background image

5

Agilent 8510SX

45 MHz to 26.5 GHz

The following specifications describe the system perfor-
mance for the 8510C network analyzer with the 8510SX
configuration. The system hardware includes the following:

Test set: 8515A S-parameter test set
RF source: 83631B synthesized sweeper
Calibration kit: 85052C 3.5mm precision
calibration kit
Calibration technique: TRL two-port calibration

Dynamic range

(for transmission measurements)

Frequency range (GHz)

0.045–2 2–8 8–20 20–26.5

Maximum power
measured at port 2

+2 dBm

+3 dBm

+3 dBm –1 dBm

Reference power
at port 1 (nominal)

–5 dBm

–9 dBm

–14 dBm –25 dBm

Minimum power
measured at port 2

–98 dBm

–98 dBm –100 dBm –99 dBm

Receiver dynamic range 100 dB

101 dB

103 dB

98 dB

System dynamic range

93 dB

89 dB

86 dB

74 dB

Measurement uncertainty

Measurement
port character-

istics

Frequency range (GHz)

RESIDUAL 0.045–2

2–8

8–20

20–26.5

Directivity

48 dB

50 dB

50 dB

50 dB

Source match

40 dB

50 dB

50 dB

50 dB

Load match

48 dB

50 dB

50 dB

50 dB

Reflection tracking

±0.003 dB ±0 dB

±0 dB

±0 dB

Transmission tracking ±0.009 dB ±0.004 dB

±0.009 dB

±0.01 dB

Crosstalk

114 dB

111 dB

106 dB

95 dB

Frequency range (GHz)

RAW (Typical)

0.045–2

2–8

8–20

20–26.5

Directivity

24 dB

24 dB

28 dB

27 dB

Source match

23 dB

23 dB

16 dB

14 dB

Load match

23 dB

23 dB

16 dB

14 dB

Reflection measurements

Transmission measurements

System dynamic range

Magnitude Phase

Magnitude Phase