Atec Agilent-8510C User Manual
Page 6

Agilent 85107B-005
45 MHz to 50 GHz
The following specifications describe the system
performance for the 8510C network analyzer in
the 85107B-005 (adds step attenuators and bias tees to
the 8517B test set) configuration for 50 GHz measure-
ments. The system hardware includes the
following:
Test set: 8517B S-parameter test set
RF source: 83651B synthesized sweeper
Calibration kit: 85056A 2.4mm calibration kit
Calibration technique: Full two-port calibration with
sliding loads
Dynamic range
(for transmission measurements)
Frequency range (GHz)
0.045–2
2–20 20–40
40–50
Maximum power
measured at port 2
+17 dBm
+8 dBm +4 dBm –3 dBm
Reference power
at port 1 (nominal)
+1 dBm
–8 dBm –19 dBm –30 dBm
Minimum power
measured at port 2
–76 dBm
–97 dBm –91 dBm –90 dBm
Receiver dynamic range 93 dB
105 dB
95 dB
87 dB
System dynamic range 77 dB
89 dB 72 dB
60 dB
Measurement uncertainty
Reflection measurements
Measurement port characteristics
Frequency range (GHz)
RESIDUAL 0.045–2 2–20
20–40 40–50
Directivity
42 dB
42 dB 38 dB 36 dB
Source match
41 dB
38 dB 33 dB 31 dB
Load match
42 dB
42 dB 38 dB 36 dB
Reflection tracking
±0.001 dB
±0.008 dB
±0.02 dB
±0.027 dB
Transmission tracking ±0.014 dB
±0.043 dB
±0.110 dB
±0.137 dB
Crosstalk
99 dB
110 dB 93 dB 81 dB
Frequency range (GHz)
RAW (Typical)
0.045–2
2–20
20–40
40–50
Directivity
22 dB
18 dB
18 dB
18 dB
Source match
20 dB
12 dB
9 dB
9 dB
Load match
20 dB
12 dB
9 dB
9 dB
Magnitude Phase
Magnitude Phase
System dynamic range