beautypg.com

Atec Agilent-8510C User Manual

Page 6

background image

Agilent 85107B-005

45 MHz to 50 GHz

The following specifications describe the system
performance for the 8510C network analyzer in
the 85107B-005 (adds step attenuators and bias tees to
the 8517B test set) configuration for 50 GHz measure-
ments. The system hardware includes the
following:

Test set: 8517B S-parameter test set
RF source: 83651B synthesized sweeper
Calibration kit: 85056A 2.4mm calibration kit
Calibration technique: Full two-port calibration with
sliding loads

Dynamic range

(for transmission measurements)

Frequency range (GHz)

0.045–2

2–20 20–40

40–50

Maximum power
measured at port 2

+17 dBm

+8 dBm +4 dBm –3 dBm

Reference power
at port 1 (nominal)

+1 dBm

–8 dBm –19 dBm –30 dBm

Minimum power
measured at port 2

–76 dBm

–97 dBm –91 dBm –90 dBm

Receiver dynamic range 93 dB

105 dB

95 dB

87 dB

System dynamic range 77 dB

89 dB 72 dB

60 dB

Measurement uncertainty

Reflection measurements

Measurement port characteristics

Frequency range (GHz)

RESIDUAL 0.045–2 2–20

20–40 40–50

Directivity

42 dB

42 dB 38 dB 36 dB

Source match

41 dB

38 dB 33 dB 31 dB

Load match

42 dB

42 dB 38 dB 36 dB

Reflection tracking

±0.001 dB

±0.008 dB

±0.02 dB

±0.027 dB

Transmission tracking ±0.014 dB

±0.043 dB

±0.110 dB

±0.137 dB

Crosstalk

99 dB

110 dB 93 dB 81 dB

Frequency range (GHz)

RAW (Typical)

0.045–2

2–20

20–40

40–50

Directivity

22 dB

18 dB

18 dB

18 dB

Source match

20 dB

12 dB

9 dB

9 dB

Load match

20 dB

12 dB

9 dB

9 dB

Magnitude Phase

Magnitude Phase

System dynamic range

6