beautypg.com

Franklin Fueling Systems TSP-LL2 User Manual

Page 8

background image

8

Troubleshooting

When troubleshooting a probe problem, it is important not

to presume any particular cause. Follow the steps outlined

below to properly and quickly troubleshoot probes:

Note: Do not start troubleshooting by swapping probes. It

may cause confusion in the process of identifying

the problem.

Float Missing:

1. Check programming of probe for correct number

of floats

2. Look for active High water or Float Missing Alarms

in the run status at the ATG.

3. If possible, have the technician stick the tank and

compare the measurement with the level reported

by the ATG.

4. Is the level correct? If not, is the ATG measuring

in reverse (the lower the actual product level the

higher it reads on the ATG)? Verify the Probe ratio

is programmed correctly. If programming is correct it

may be a probe issue, continue with the next steps.

5. Check the water level does the ATG list it as float

missing? If it is listed as missing, it could be a

probe issue, but again continue to troubleshoot.

6. Try to cycle power. If previously noted, check to

see if High water alarm or float missing alarm

changed from “ACTIVE” to “CLEARED”. If they

clear, again this could be a probe issue, but

acknowledge and continue to troubleshoot.

7. After completing the above steps, replace the

probe.

No Probe, Probe Sync and Unstable Probe

(New Install or Replacement Probe)

1. Check Wiring. Ensure that the Red, Black and

Shield wires are connected to the correct terminals

to the ATG. There should be two connections in

the junction box, (refer to figure 8).

(Probe Alarms on Multiple Channels or Only One Probe

Connected)

2. Check Main Board Power Supply. There should be

approximately 21 VDC loaded (probe connected)

and 24 VDC unloaded across the Red and Black

terminals of the probe channel in the ATG.

3. Check Field Wiring. Ensure that the voltage is

getting to the probe by checking at the junction box

splice or quick disconnect pigtail. With the pigtail

end facing you, keyslot up, there should be 24

VDC across the two bottom pins. Test field wiring

for a short or open.

4. Remake Splice. Cutback and remake the splice

connection in the junction box using the approved

seal pack and connectors.

(Quick Disconnect Pigtail)

5. Try a New Pigtail. If this probe has a quick

disconnect pigtail, try replacing it.

6. Try another Probe Channel. Once the field wiring

has been checked and the junction box connection

remade, try moving the field wiring to another

channel.

7. Connect the Probe Directly to the ATG. If possible,

bring the probe inside and wire directly to the ATG.

(If this probe has a quick disconnect pigtail then

the white wire should be connected to shield or

ground).

Listed below are additional steps for troubleshooting

the probe. For missing float or incorrect level problems,

consider the following actions:
• Pull probe from tank and manually raise and lower floats.

Does the ATG accurately report the change in level?

• Are the right number of floats installed on the probe

(verify number of floats programming)

• Is float too close to the end of the probe? (Was tank

overfilled or is the float stuck in riser)? If float is too close

to the probe head a float missing alarm will occur.

• Is there debris on the magnets? If so, clean debris from

magnet and check operation of ATG.

• Are the magnets cracked or broken? If so replace float
• Are the floats installed correctly? Product float on top,

closest to the probe head. Product float will have a white

collar. Water floats have colored collars (Red for use in

gasoline, blue for use in diesel)

• Is the probe shaft clean? If dirty use an abrasive pad to

clean the probe shaft.

• Check that wire connections from probe cable to field

wiring are good and dry.

• Are the floats compatible with the probe? LL and LL2

floats are NOT interchangeable.