beautypg.com

4 relay output r1/r2/r3, ma output c1/c2, 5 system, Relay output r1/r2/r3, ma output c1/c2 -4 – YSI IQ S ENSOR N ET System 182 User Manual

Page 88: System -4

background image

The SETTINGS menu

System 182

5 - 4

ba76022e01

12/2011

5.4

Relay output R1/R2/R3, mA output C1/C2

In both menus you can link sensors with the relay and current outputs
of the Universal Transmitter and adjust the outputs. The various setting
options are described in detail in the chapter 6 O

UTPUTS

menu.

5.5

System

The settings in the SYSTEM menu comprise:

Date/Time

Pressure/Altitude

This setting is important if you want to carry out D. O. measurements
with the System 182.
(adjustable range: 500 ... 1100 mbar or 0 ... 5955 m altitude).

Settings

a

/

z?

Here you can configure the password for the settings.
For details on the passwords refer to section 4.6.

Calibrate

a

/

z?

Here you can configure the password for calibration.
For details on the passwords refer to section 4.6.

Display contrast

Here you can set the display contrast for optimum readability.

Delivery state

Here you can reset all settings to default. Only the date, time and
contrast setting for the display will remain stored. If the settings
password is active, a password prompt appears before the reset. In
any case a security prompt appears to prevent an inadvertent reset.