Outputs
System 182
6 - 24
ba76022e01
12/2011
Canceling the cleaning
A running cleaning cycle is canceled:
Automatically
– If the sensor switches to the inactive condition during the cleaning
cycle
Manually
– By pressing
c
– By switching on the maintenance condition
Each time the cleaning cycle is canceled, the relay opens immediately.
If the cleaning cycle is canceled automatically, the outputs linked to the
sensor are released immediately.
If the cleaning cycle is canceled manually, the sensor is in the
maintenance condition. The linked outputs are only released after the
maintenance condition was terminated manually.
The next cleaning cycle will be performed at the time set up.
Note
In case of a power failure, all relays open. The cleaning cycle is
canceled. The outputs linked with the sensor change to the non-active
condition (see section 6.6.2). As soon as the power is available, the
outputs are released again. The next cleaning cycle will be performed
at the time set up.
6.4.9
Manual control
Function
With the Manual control function, you can test the operativeness of an
instrument that is connected to the relay. To do so, close or open the
relay manually and, while doing so, check the behavior of the
connected instrument.
Note
The settings for other functions in the Relay function menu, as for
example Frequency controller and Pulse-width output, are retained
while the Manual control is carried out.
Setting
Selection/Values
Explanation
Relay function
Manual control
The selected relay action
is carried out with Save
and quit.
Action
Open
Close
Relay action