beautypg.com

Warning – YSI IQ S ENSOR N ET System 182 User Manual

Page 60

background image

Installation

System 182

3 - 46

ba76022e01

12/2011

3.11.2 Connecting two sensors with compressed air cleaning

Connection scheme of

valve control

Example:
Relay 1 controls the cleaning of sensor 1.
Relay 2 controls the cleaning of sensor 2.

Variant: Relay 1 controls the cleaning of both sensors. Cleaning of both
sensors is carried out with the same settings.

Warning

In this configuration, the free switching contact (here: R3) may be
used to switch SELV voltages only.

Warning

No free wires must be allowed to project into the enclosure.
Otherwise, there is a danger that areas safe to contact could come
into contact with dangerous voltages. Always cut off any wires
that are not in use as closely as possible to the cable gland.

M

C

S

ESC

OK

DIQ/CHV 1

DIQ/JB

SNCIQ(/UG)

SACIQ

SACIQ

Valve control line 1

Valve control line 2

Air

Air

DIQ/CHV 2

Sensor 2

Sensor 1

Maximum total cable length
SNCIQ(/UG) plus SACIQ = 250 m

DIQ/S 182(/24V)