YSI IQ S ENSOR N ET System 182 User Manual
Page 69
System 182
Operation
4 - 3
ba76022e01
12/2011
Select the PROPERTIES menu with
g
. With this menu you can put
sensors in the maintenance condition and prompt important data
(calibration data, error messages, operating states, software
versions, etc.).
Open the SETTINGS menu with
s
.
Special
sensor conditions
The following displays inform you of special states of sensors
Error and info symbol
If the info symbol
i
or error symbol
h
appears in the header, error
messages from the sensors or error conditions of the system have
occurred. Error messages from the sensors are entered in the log book.
The log book is described in detail in section 4.4.3. Error conditions of
the system can e.g. be an insufficient operational voltage or
malfunctions of the communication and are displayed in the
PROPERTIES / SYSTEM STATUS menu (see section 4.4.1).
Init
Sensor is being initialized
during commissioning or
if a new IQ sensor is recognized that is not yet
giving measured values
----
Sensor inactive or inadmissible operating conditions
Cal
Sensor is being calibrated
Clean
Cleaning procedure active
Error
No communication with the sensor due to defective
connection
OFL
Measuring range undercut or exceeded (overflow)
Display
flashes
Sensor in maintenance condition