Electrical data iq s, 24v (24 v ad/dc version), Electrical connections iq s – YSI IQ S ENSOR N ET System 182 User Manual
Page 133: Ensor
System 182
Technical data
9 - 3
ba76022e01
12/2011
Electrical data
IQ S
ENSOR
N
ET
/24V
(24 V AD/DC version)
Electrical connections
IQ S
ENSOR
N
ET
(line power version)
Terminal strip inside the enclosure:
Electrical connections
IQ S
ENSOR
N
ET
/24V
(24 V AC/DC version)
Terminal strip inside the enclosure:
Supply
Nominal voltage:
24 V AC/DC ± 10 %
protective low voltage
SELV (Safety
Extra Low Voltage)
AC frequency:
50/60 Hz according to
DIN IEC 60038
Connection:
2 pin
Line cross-section of connections:
Europe:
1.5 ... 4.0 mm
2
USA:
AWG 14 ... 12
Fuse rating on the operator side: max. 16 A
Power consumption
max. approx. 12 W
SENSORNET 1
RED
SHIELD
GREEN
X3
X2
X17
X9
X15
X13
X7
X11
X5
X1
X16
X8
X14
X12
X6
X10
X4
100...
240V AC
MAINS
AUXILIARY
R3
R2
R1
L1
N
240V AC
2A AC
≤
≤
240V AC
2A AC
≤
≤
240V AC
2A AC
≤
≤
AUXILIARY
VOLTAGE
0/4...20mA
+ REC -
C2
0/4...20mA
+ REC -
C1
SENSORNET 1
RED
SHIELD
GREEN
X3
X2
X17
X9
X1
X16
X8
24V AC DC
INPUT
POWER
AUXILIARY
AUXILIARY
VOLTAGE
X7
X5
X6
X4
0/4...20mA
+ REC -
C2
0/4...20mA
+ REC -
C1
X15
X13
X11
X14
X12
X10
R3
R2
R1
240V AC
2A AC
≤
≤
240V AC
2A AC
≤
≤
240V AC
2A AC
≤
≤