5 frequency controller, Freque, Section 6.4.5 – YSI IQ S ENSOR N ET System 182 User Manual
Page 108
Outputs
System 182
6 - 18
ba76022e01
12/2011
6.4.5
Frequency controller
Function
The characteristic of the frequency output is laid down in the Start
value, End value, Frequency (f) min. and Frequency (f) max. settings.
The fundamentals of the function are described in the introductory
chapter (see section 6.3.3).
In order to set up the Frequency controller function for a relay output,
the relay output must be linked with a sensor (see section 6.2.1).
Settings
Characteristic curve
If a value is entered for End value that is greater than the Start value,
the output has a positive characteristic curve.
In order to obtain a negative characteristic curve, a value must be
entered for End value that is smaller than the value for Start value.
Setting
Selection/Values
Explanation
Measured variable
Main variable
Adjoining variable
Main variable
designates the
actual measured
parameter of the
sensor (e.g. pH,
oxygen, etc.).
Adjoining variable
designates an
additional
measured
parameter (e.g.
temperature).
Start value
within the
measuring range
(sensor-
dependent)
Minimum spacing:
5 % of the
measuring range
End value
Frequency (f) min.
0 to 120 1/min
Minimum spacing:
10 1/min
Frequency (f) max.
Frequency with error
0 to 120 1/min
In case of system
errors or sensor
errors (see page 6-
33), the relay
switches with the
frequency
specified.
Action
Open
Close
Relay action