4.4.2

Maintenance condition

When an IQ sensor is calibrated, cleaned, serviced or repaired, the
maintenance condition for the relevant IQ sensor should always be
switched on.

In the maintenance condition

 apart from the measured value display, the system does not react to

the current measured value or the condition of the selected
IQ sensor

 IQ sensor errors do not prompt changes in the conditions of linked

outputs.

The following diagram gives you an overview of when an IQ sensor is
in the maintenance condition.

= Maintenance condition ON

g m e

/

/

m

c

e/m

g

c

g

Measured value

display

The measured value

stops flashing

Measured value

display

The measured

value flashes

Measured value

display

flashes

Clean

Calibration

Step 1

Calibration

Step n

End

Maintanance

condition ON

manually

Maintanance

condition OFF

manually

Sensor

cleaning

active

Sensor

cleaning

OFF

(automatic sequence)

= Maintenance condition OFF

e

Switch

Display message:

on the

maintenance

condition

Continue/Cancel