Setting logging interval, Storing barometer readings – YSI 556 MPS User Manual
Page 73
Logging
Section
9
YSI Incorporated
YSI 556 MPS
Page 67
9.2 Setting Logging Interval
Follow steps below to set the interval for logging a data stream.
NOTE: If you do not specify an interval, the instrument will use a default
interval setting of 1 second.
NOTE: It is not necessary to set a logging interval when logging a single
sample.
1. Go to the logging setup screen as described in Section 9.1
Accessing the Logging Setup Screen.
2. Use the keypad to enter an interval between 1 second and 15
minutes. Refer to Section 2.9 Keypad Use.
NOTE: The interval field has hour, minute and second entry
fields. Any entry over 1 hour will change automatically to a
15-minute setting.
3. Press the Enter key. The data stream interval is set.
4. Press the Escape key repeatedly to return to the main menu
screen.
9.3 Storing Barometer Readings
NOTE: The Store barometer option is only available on
instruments that are equipped with the optional barometer.
1. Go to the logging setup screen as described in Section 9.1
Accessing the Logging Setup Screen.
2. Use the arrow keys to highlight the Store barometer
selection. See Figure 9.2 Logging Setup Screen.
3. Press the Enter key until a check mark is entered in the box
next to the store barometer selection if you want to log
barometric readings.