YSI 556 MPS User Manual
Page 49

Calibrate
Section
6
YSI Incorporated
YSI 556 MPS
Page 43
9. Observe the reading under DO %. When the reading shows no
significant change for approximately 30 seconds, press Enter.
The screen will indicate that the calibration has been accepted
and prompt you to press Enter again to Continue. See Figure
6.6 Calibrated.
10. Press Enter. This returns you to the DO calibration screen,
See Figure 6.7 DO Calibration Screen.
11. Press Escape to return to the calibrate menu. See Figure 6.2
Calibrate Screen.
12. Rinse the probe module and sensors in tap or purified water
and dry.
DO Calibration in mg/L
DO calibration in mg/L is carried out in a water sample which has a known
concentration of dissolved oxygen (usually determined by a Winkler
titration).
1. Go to the DO calibrate screen as described in Section 6.2.3
Dissolved Oxygen Calibration, steps 1 through 3.
2. Use the arrow keys to highlight the DO mg/L selection.
3. Press Enter. The DO mg/L Entry Screen is displayed.
- 3100 (50 pages)
- 3200 (86 pages)
- 5000 (70 pages)
- 5200 (148 pages)
- 5200A (128 pages)
- 5200A (121 pages)
- 5400 (127 pages)
- 5239 Dissolved Oxygen Probe (23 pages)
- 55 (26 pages)
- 550A (18 pages)
- 556 MPS - Quick Start Guide (4 pages)
- 58 (40 pages)
- 5905 (22 pages)
- 6600V2 - User Manual (379 pages)
- 60 (33 pages)
- 63 (46 pages)
- 650 MDS (72 pages)
- 9300 Photometer (11 pages)
- 9500 Photometer (143 pages)
- 9500 Photometer (18 pages)
- CR 3200 Thermoreactor (42 pages)
- CR 4200 Thermoreactor (42 pages)
- CR2200 Thermoreactor (36 pages)
- DO200 (40 pages)
- EC300 (40 pages)
- EC300A (40 pages)
- EC30A (12 pages)
- IFL 700 IQ (46 pages)
- IQ SensorNet 2020 XT Terminal (196 pages)
- IQ SensorNet FDO Sensor (48 pages)
- IQ SensorNet MIQ/A-MOD Module (42 pages)
- IQ SensorNet MIQ/A-PR Module (36 pages)
- IQ SensorNet MIQ/IC2 Module (30 pages)
- IQ SensorNet MIQ/IF232 Module (18 pages)
- IQ SensorNet MIQ/JB Module (18 pages)
- IQ SensorNet MIQ/JBR Module (20 pages)
- IQ SensorNet MIQ/PS Module (22 pages)
- IQ SensorNet MIQ/R6 Module (56 pages)
- IQ SensorNet NitraLyt Sensor (64 pages)
- IQ SensorNet SensoLyt Sensor 700 IQ (SW) (46 pages)
- IQ SensorNet Sensors Encoded Fieldbus Data (12 pages)
- IQ SensorNet TetraCon Sensor 700 IQ (SW) (32 pages)
- IQ SensorNet ViSolid Sensor 700 IQ (60 pages)
- MultiLab 4010-1 (108 pages)
- MultiLab 4010-2 (138 pages)