Algorithms used, Appendix a, Ph calculation – Ocean Optics pH Sensor Patches, Probes and Cuvettes User Manual

Page 31: Temperature compensation, Resetting pk and slope

background image

PHSENSOR-TRANS-REFLECT-201111

25

Appendix A

Algorithms Used

pH Calculation



Sample

pH

Sample

Abs

Abs

Abs

Slope

pK

pH

11

log

*

…where Abs

Sample

is the sample absorbance at 620nm with baseline correction, and Abs

pH11

is

the absorbance at pH 11 at 620nm with baseline correction.

Temperature Compensation

When you select Use Factory Calibration in SpectraSuite, the value for pK is adjusted via the
van’t Hoff equation based on the current temperature you entered:









1

2

1

2

1

1

*

480

1

2

1

1

*

480

1

2

log

log

T

T

T

T

e

pH

pH

e

pK

pK

Resetting pK and Slope

An x-y plot is made using data obtained from intermediate buffers 5 through 8. The x-axis is of
the term:



Sample

pH

Sample

Abs

Abs

Abs

11

log