Jaz software installation and operation, Performing a factory calibration with jaz – Ocean Optics pH Sensor Patches, Probes and Cuvettes User Manual

Page 23

background image

2: Desktop Software Installation and Operation

PHSENSOR-TRANS-REFLECT-201111

17

3. Click one or all of the following buttons:

Reset: Clears the table and restarts the run time.
Export: Opens a window to save your data in a format that can be opened with Microsoft

Excel or a text program such as WordPad. The exported data file contains all of the variables
that you have entered and have been calculated, along with a time stamp for data acquisition
and save, the time-resolved pH data, and complete spectra for all reference and calibration
buffers used.

Export Calibration: Opens a window to save your calibration data. This creates a file

containing the reference spectra and other variables that can later be loaded via the
Calibration Wizard, allowing for very quick setup.

Jaz Software Installation and Operation

The Jaz pH Application should be loaded on an SD card that is inserted in the Jaz when the Jaz is
powered up. This will cause the application to run automatically.

Performing a Factory Calibration with Jaz

Procedure

1. Select the Setup/Calibrate option from the Jaz screen.