Taking ph measurements with jaz – Ocean Optics pH Sensor Patches, Probes and Cuvettes User Manual
Page 28

2: Desktop Software Installation and Operation
22
PHSENSOR-TRANS-REFLECT-201111
10. Take a high reference spectrum for pH 11.0. To do this, expose the sensor to pH 11 buffer in the
manner described previously in the SpectraSuite operating instructions, based on your sensor form
factor. Wait the appropriate time for equilibration. When complete, select the Get pH 11 Ref option
on the Jaz.
11. Repeat step 10 for pH buffers 5, 6, 7, and 8, as they are requested by the software.
12. When these steps are completed a similar screen with your parameters will display for 5 seconds.
You are now ready to take pH measurements.
Taking pH Measurements with Jaz
Now that you have finished calibration, you can take pH measurements in the biological range.
There are three different ways to take pH measurements.
Take a single pH measurement and display it to the Jaz screen.
Take continuous pH measurements and display them to the Jaz screen.
Take continuous pH measurements, display them to the Jaz screen and write them to a data
file on an SD card in the Jaz.