Appendix A

Saving the New Calibration Coefficients: Serial Mode

If you are connecting the HR2000 Spectrometer to the serial port of the PC, you need to save the new wavelength
calibration coefficients to the .SPEC file that OOIBase32 accesses when opened.

Note: You cannot save the calibration coefficients to the EEPROM memory chip on the HR2000 when using the

serial mode.

To Save Wavelength Calibration Coefficients Using the Serial Mode

1. Open the OOIBase32 application.

2. Select

Spectrometer | Configure from the OOIBase32 menu bar. The Configure Spectrometer screen

appears.

3. Select

the

Wavelength Calibration tab to update the wavelength coefficients within OOIBase32.

4. Enter in the new values acquired from Steps 5 and 6 of the Calibrating the Wavelength of the

Spectrometer section in this Appendix.

HR2000 High-Resolution Fiber Optic Spectrometer

35

5. Click

the

OK button to save the information in OOIBase32.