beautypg.com

Preparing for experiments, Chapter, Sample experiments – Ocean Optics HR2000CG-UV-NIR User Manual

Page 22: 5 sample experiments

background image

Sample Experiments

HR2000 High-Resolution Fiber Optic Spectrometer

18

5 Sample

Experiments

The following sections contain information on conducting sample experiments using the HR2000 and OOIBase32.
For information on experiments with OOIPS2000, consult the OOIPS2000 Operating Instructions.

Preparing for Experiments

Follow the steps below to configure the HR2000 and OOIBase32 for experiments:

1. Double-check that you have correctly installed the HR2000, installed OOIBase32, and configured the light

source and other sampling optics.

2. Open the OOIBase32 application, select Spectrometer | Configure from the menu bar, and double-

check that A/D Interface settings are correct.

3. Check your spectrometer setup configurations in OOIBase32:


Locate the Wavelength Calibration Data sheet that came with the HR2000. Select Spectrometer |
Configure
from the menu and choose the Wavelength Calibration page. For each spectrometer
channel in the system, enable the channel and make sure the First Coefficient, Second Coefficient, Third
Coefficient and Intercept correspond to those of the system.

4. Adjust the acquisition parameters using the Acquisition Parameters dialog bar or select Spectrum |

Configure Data Acquisition from the menu.

If you have followed the previous steps and started OOIBase32, the spectrometer is already acquiring
data. Even with no light in the spectrometer, there should be a dynamic trace displayed in the bottom of
the graph. If you allow light into the spectrometer, the graph trace should rise with increasing light
intensity. This means the software and hardware are correctly installed.

Once you install the hardware, configure the software, and establish your sampling system, you are ready to take
measurements.

There are four basic optical measurements from which to choose:

• Absorbance
• Transmission
• Reflection
• Relative

irradiance

The type of measurement you will take determines the configuration of the sampling optics for your system.
Furthermore, your choice of reference and data analysis determines how the OOIBase32 presents the results.

Note: For each measurement, you must first take a reference and dark spectrum. After you take a reference

and a dark spectrum, you can take as many measurement scans as needed. However, if you change any
sampling variable (integration time, averaging, smoothing, angle, temperature, fiber size, etc.), you must
store a new dark and reference spectrum.