beautypg.com

Hr2000 specifications, Ccd detector specifications, Hr2000 spectrometer specifications – Ocean Optics HR2000CG-UV-NIR User Manual

Page 11

background image

HR2000 Specifications

HR2000 High-Resolution Fiber Optic Spectrometer

7

HR2000 Specifications

The following sections provide specification information for the CCD detector in the HR2000, as well as the
HR2000 Spectrometer itself.

CCD Detector Specifications

Detector:

Sony ILX511 linear silicon CCD array

No. of elements:

2048 pixels

Pixel size:

14 µm x 200 µm

Pixel well depth:

62,500 electrons

Signal-to-noise ratio:

250:1 (at full signal)

A/D resolution:

12 bit

Dark noise:

2.5 RMS counts

Corrected linearity:

>99.8%

HR2000 Spectrometer Specifications

Dimensions:

148.6 mm x 104.8 mm x 45.1 mm

Weight: 570

g

Power consumption:

95 mA @ 5 VDC

Detector range:

200-1100 nm

Gratings:

13 gratings available

Entrance aperture:

5, 10, 25, 50, 100 or 200 µm wide slits or fiber (no slit)

Order-sorting filters:

Installed longpass and bandpass filters

Focal length:

f/4, 101 mm

Optical resolution:

Depends on grating and size of entrance aperture

Stray light:

<0.05% at 600 nm; <0.10% at 435 nm

Dynamic range:

2 x 10

8

(system); 2000:1 for a single scan

Fiber optic connector:

SMA 905 to single-strand optical fiber (0.22 NA)

Data transfer rate:

Full scans into memory every 13 milliseconds

Integration time:

3 milliseconds to 65 seconds

Operating systems:

Windows 98/Me/2000/XP when using the USB interface on desktop or
notebook PCs
Any 32-bit Windows operating system when using the serial port on
desktop or notebook PCs