Automatic calibration printout, User calibration printout, Calibration test printout – Ohaus AP250E ANALYTICAL Plus Electronic Balances User Manual

Page 29: Operation

background image

29

Printing Data (Cont.)

Printing Data (Cont.)

Printing Data (Cont.)

Printing Data (Cont.)

Printing Data (Cont.)

Automatic Calibration Printout

Automatic Calibration Printout

Automatic Calibration Printout

Automatic Calibration Printout

Automatic Calibration Printout

When performing an Automatic calibra-
tion with CAL option (GLP submenu of the
Setup Options submenu set to ON), a
printout is made after calibration is com-
pleted.

Calibration Test Printout

Calibration Test Printout

Calibration Test Printout

Calibration Test Printout

Calibration Test Printout

When performing a Calibration Test with
with CAL option (GLP submenu of the
Setup Options submenu set to ON), a
printout is available.

User Calibration Printout

User Calibration Printout

User Calibration Printout

User Calibration Printout

User Calibration Printout

When performing User calibration with
CAL option (GLP submenu of the Setup
Options submenu set to ON), a printout is
made after calibration is completed.

- - - - - AUTO SPAN CAL - - - - - -
6/22/95

8:42:24 AM

Bal Id 183
Auto. Cal. completed !
Dif:

- 0.00136g

ID 2000000
PR 10000
Name........................................

- - - - - END - - - - -

- - - - - USER SPAN CAL - - - - - -
6/22/95

8:52:21 AM

Bal Id 183
Cal:

200.0000g

Old:

200.0398g

Dif:

0.0398g

Wt. Ref......................................
ID 2000000
PR 10000
Name........................................

- - - - - END - - - - -

- - - - - CAL TEST - - - - - -
6/22/95

8:47:02 AM

Bal Id 183
Cal. test completed !
Dif:

-0.00045g

ID 2000000
PR 10000
Name........................................

- - - - - END - - - - -

OPERATION