Installation – Ohaus AP250E ANALYTICAL Plus Electronic Balances User Manual
Page 13

13
Command
Command
Command
Command
Command
Character
Character
Character
Character
Character
Description
Description
Description
Description
Description
xS
xS
xS
xS
xS
Set stable data only printing (set x = 0 for OFF, or x = 1 for ON.
T
T
T
T
T
Same effect as pressing on tare button.
V
V
V
V
V
Print EPROM version
x#
x#
x#
x#
x#
Set current ref., any function, CW takes two reference separated by
a space.
x%
x%
x%
x%
x%
Doanloads reference weight "x" for percent mode. "x" must be in
grams. Command is ignored if percent mode is disabled. If percent
mode is enabled, balance will automatically switch to percent mode
display.
xF
xF
xF
xF
xF
Set current function to "x". x = 0 to 6. Setup menu must be unlocked.
0 = None
1 = Percent
2 = Parts Counting
3 = Check Weighing
4 = Animal Weighing
5 = Fill Guide
6 = High Point
xZ
xZ
xZ
xZ
xZ
Set Auto Zero to "x". x = 0 for OFF, x = 1 for ON
xSL
xSL
xSL
xSL
xSL
Set stability level. User menu must be unlocked. x = 0 to 3
Z
Z
Z
Z
Z
Zero request (Gross tare) if Net/Gross enabled.
nnnA
nnnA
nnnA
nnnA
nnnA
Set Auto Print feature to "nnn"(see table).
nnn = 0
Turns feature OFF
nnn = S
Output on stability
nnn = C
Output is continuous
nnn = 1-256
Sets Auto Print interval
Esc L
Esc L
Esc L
Esc L
Esc L
Prints listing of Setup and Print menu settings.
Esc R
Esc R
Esc R
Esc R
Esc R
Resets Setup and Print menus to factory defaults.
CAUTION: This will reset RS232 configuration.
Esc S
Esc S
Esc S
Esc S
Esc S
Save current settings.
Field: Model # 1 EPROM #
CR
LF
Length:
6
16
1
1
Balance Model
"98101-xx Sr#x.xx"
INSTALLATION