Printing data, Time and date, Operation – Ohaus AP250E ANALYTICAL Plus Electronic Balances User Manual
Page 27

27
NOT DISPLAYED
5g
5g
5g
5g
5g
4g
4g
4g
4g
4g
2g
2g
2g
2g
2g
20g
20g
20g
20g
20g
40g
40g
40g
40g
40g
ORDER SAMPLES TAKEN IN
5. To use the High Point function again, repeat steps 1 through 4.
6.
to exit High Point and return to a weighing
mode.
High Point (Cont.)
High Point (Cont.)
High Point (Cont.)
High Point (Cont.)
High Point (Cont.)
4.
To view the lowest and highest sample weight. The display LIMIT
flashes, the lowest sample weight is displayed followed by two short beeps,
the
display then indicates the highest sample weight for a few seconds then
automatically changes back to the normal weighing mode.
6/22/95
1:00:30 PM
Printing Data
Printing Data
Printing Data
Printing Data
Printing Data
Printing data to an external computer or printer requires that the communications
parameters in the Print menu be set first. Refer to page 51 Print menu. A wide variety
of printing options are available, refer to page 55, Print Options under the Print menu
and set the desired options before proceeding. To print data,
.
This section defines the various printing setups with printing samples.
Time and Date
Time and Date
Time and Date
Time and Date
Time and Date
When time and date are entered in the
balance through the Setup menu and with
both Time and Date options set to ON
under the Print Options submenu, each
printout starts with the time and date on the
first line.
OPERATION