5.4
Technical specifications and illustrations are without
obligation, subject to modifications. 08.99/06.06
M&C-Dilution unit DIL-1/ (H)
Dilution rates with the critical orifices ”a“ - ”g“
a = 500
b = 200
c = 100
d = 50
e = 30*
f = 20
g = 10
: 1
Sample flow rate depending on the critical orifices
”a“- ”g“
a = 1,4
b = 2,7
c = 5,5
d = 11
e = 19*
f = 28
g = 55
l/hr
1)
Possibility to adapt the dilution factor
with dilution gas pressure-adjustment -5% bis +30%
2)
Dilution gas flow rate with injector version I or II
I: 480 - 600 Nl/hr, optional II: 1800 – 3000 Nl/hr
Dilution gas pressure on inlet of pressure controller
min. 4,5 bar g, max. 16 bar g
Bypass injector / B sample gas flow rate
gas pressure / gas flow rate
approx. 2 bar
injector gas approx. 300 l/hr sample gas approx. 150 l/hr
Process pressure
0,9 to 2 bar abs.
Fault caused by process temperature variations
operation independent of process temperature
Fault caused by process low- or overpressure
no fault as long as the differential pressure ∆P at the dilution unit is > 0,5 bar g and test gas is given to
the probe under process conditions
Fault caused by atmospheric pressure variations
< 1% with a variation of 50 mbar
Materials in contact with the sample gas
stainless steel 316 / 316Ti, quartz glass, FPM, graphite
Power supply for DIL-1/H
230V 50Hz, 800W
Temperature controller for DIL-1/H
capillary thermostat adjustable 0-180 °C , with high temperature limiter
and low temperature alarm as changeover contact, voltage free
alarm point ∆T30 °C to TSet, contact rating max. 250V 3A~ 0,25A =
Weight
approx. 8 kg
Technical Data
Part-No.
Type
M&C-Dilution Unit DIL-1/ (H) with orifice ”e“ for dilution ratio 30 : 1 standard
20 S 4900
DIL-1
Dilution unit non heated
20 S 4905
DIL-1/H (a)
Dilution unit electrically heated to 180 °C , (a) to be added to the Part-No. for power 115V 60Hz
20 S 4925
Option I
Bypass injector - B
20 S 4930
Option II
Additional sample gas outlet undiluted
20 S 4205
DIL-1/-A
option: 1 pressure controller, 2 gauges, mounting set for dilution unit
20 S 4215
DIL-1/-A1
option: 2 pressure controllers, 3 gauges, mounting set incl. connector set for dilution unit with by-pass
20 S 4250
DIL-1/-S
option: mounting set with 1 pressure controller, 2 gauges, flowmeter, 2 ball valves
20 S 4260
DIL-1/-S1
option: mounting set with 2 pressure controllers, 3 gauges, flowmeter, 2 ball valve
Other executions on request.
* Standard, others to be indicated along with order, intermediate values possible.
1) approx. at 3 bar dilution gas.
2) -5% not possible for orifice ”g“.