beautypg.com

M&C TechGroup CLF-SS Operator's manual User Manual

Fluid particle filter series clf, Special features, Application

background image

M&C TechGroup Germany GmbH • Rehhecke 79 • 40885 Ratingen • Germany

[email protected] • www.mc-techgroup.com • Fon +49 2102 935-0 • Fax +49 2102 935-111

7.8

Embracing Challenge

ƒ

ƒ

High retention rate of 99,99% for parti-
cles > 0,1µm

ƒ

ƒ

High chemical resistance of type CLF-T

ƒ

ƒ

Type CLF-SS suitable for high pressure
applications

ƒ

ƒ

Condition of filter element CLF-T visible
from outside

ƒ

ƒ

Easy change of filter element

ƒ

ƒ

Wall mounting

Special Features

Fluid Particle Filter Series CLF®

Version CLF-T, CLF-SS made out of special materials,
for removal of aerosols from gases

CLF-T

Application

The M&C fluid particle filter CLF-.. is suitable

for filtration of fluid particles of all types and

is recommended for sample gases with an

acid dew point above 100 °C. Examples are

measurements in flue gas of heavy oil and

black coal combustions.
The filter separates the aerosols (very fine fluid

particles) which still pass the gas cooler. The

most effective position of the CLF-.. filter is

downstream the sample conditioning close

to the flowmeter of the analyser.

Description

The filter element of the M&C fluid particle

filter is constructed in two sections with a

flow direction from the inside to the outside

of the filter element. The inner, very fine, fibre

layer binds the fluid particles suspended in

the gas and leads them together with the gas

flow to the outer, larger layer. On their way

through the filter element, the very fine fluid

particles accumulate with others and form

droplets. The vertical flow direction and the

force of gravity cause the droplets to drip into

the filter pot.
The filter element remains fully effective even

when completely saturated with fluid. If it is

not effected by solid particles, the lifetime

is nearly unlimited. The compressed micro-

fibres are made with binding of PVDF in order

to prevent influences on the sample gas.
The condition of the CLF-T filter is visible

through the glass body without opening the

filter.

The separated acid mist can continuously be

discharged with an external mounted peri-

staltic pump SR25.1 or a liquid drain AD-SS

(option) connected by a GL25 or a G3/8“

adapter.
No tools are required to change the filter ele-

ment. The optimised position of the O-ring

always guarantees a safe sealing of the filter

body to the filter head. The filter in- and out-

let can be turned by 180 ° on the mounting

bracket to achieve easy mounting and flexible

adaption to local circumstances.
The filter is made out of PTFE/glass or stainless

steel 316Ti.

Handle with care! Danger of acid burns!

Technical specifications and illustrations are without

obligation, subject to modifications. 06.99/06.06