11.4

Embracing Challenge

ƒ

ƒ

Corrosion resistant

ƒ

ƒ

Compact design

ƒ

ƒ

Gas tight

ƒ

ƒ

Maintenance free

ƒ

ƒ

Easy to operate

ƒ

ƒ

Easy mounting

ƒ

ƒ

Controlled position as option

Special Features

Electrically heated connectors
series T..-H1, V..-H1

Version T6-H1, T8-H1, tee tube union, electrically
heated to 180 °C
Version V6-H1, V8-H1, straight tube union, electrically
heated to 180 °C

T6-H1

Application

In analysis technique, often temperatures

must be kept above the sample gas dew

point. Therefore, it is absolutely necessary to

avoid cold bridges.
In order to ensure this, the temperature regu-

lated M&C hose-/tube connectors T..-H1 and

V..-H1 are used to connect heated sample

lines type 3/4/5-N/M/H with an operating

temperature up to 180 °C.

Description

The M&C hose-/tube connectors T..-H1 and

V..-H1 are fixed on a mounting plate, decou-

pled from heat and covered with an insulated

enclosure.
The heater consists of a heating element

with high capacity. The temperature is adjust-

able on the integrated thermostat up to 180

°C with high temperature limiter and low

temperature alarm. The connection box with

integrated thermostat is installed outside the

enclosure on the mounting plate.
The bushings for the heated sample lines in

the enclosure are insulated with sockets of

heat resistant silicone. In order to avoid cold

bridges, the connector is completely heated

by means of a double-ended thermal con-

ducting jaw. For fixing the electrically heated

sample lines type 3/4/5-N/M/H – see data

sheet 3.1 –, mounting brackets are integrated.

Technical specifications and illustrations are without

obligation, subject to modifications. 08.98/11.06