M&C TechGroup V..-H1 Series Data sheet User Manual
Special features, Application, Description
M&C TechGroup Germany GmbH • Rehhecke 79 • 40885 Ratingen • Germany
[email protected] • www.mc-techgroup.com • Fon +49 2102 935-0 • Fax +49 2102 935-111
11.4
Embracing Challenge
Corrosion resistant
Compact design
Gas tight
Maintenance free
Easy to operate
Easy mounting
Controlled position as option
Special Features
Electrically heated connectors
series T..-H1, V..-H1
Version T6-H1, T8-H1, tee tube union, electrically
heated to 180 °C
Version V6-H1, V8-H1, straight tube union, electrically
heated to 180 °C
T6-H1
Application
In analysis technique, often temperatures
must be kept above the sample gas dew
point. Therefore, it is absolutely necessary to
avoid cold bridges.
In order to ensure this, the temperature regu-
lated M&C hose-/tube connectors T..-H1 and
V..-H1 are used to connect heated sample
lines type 3/4/5-N/M/H with an operating
temperature up to 180 °C.
Description
The M&C hose-/tube connectors T..-H1 and
V..-H1 are fixed on a mounting plate, decou-
pled from heat and covered with an insulated
enclosure.
The heater consists of a heating element
with high capacity. The temperature is adjust-
able on the integrated thermostat up to 180
°C with high temperature limiter and low
temperature alarm. The connection box with
integrated thermostat is installed outside the
enclosure on the mounting plate.
The bushings for the heated sample lines in
the enclosure are insulated with sockets of
heat resistant silicone. In order to avoid cold
bridges, the connector is completely heated
by means of a double-ended thermal con-
ducting jaw. For fixing the electrically heated
sample lines type 3/4/5-N/M/H – see data
sheet 3.1 –, mounting brackets are integrated.
Technical specifications and illustrations are without
obligation, subject to modifications. 08.98/11.06